Diagnostic support, emergency prioritisation, imaging analysis, prediction of decompensation, prescribing assistants: AI in healthcare has moved from the laboratory to clinical routine. Legally, it is the densest use case there is: a single piece of software can fall simultaneously under the Medical Devices Regulation, the AI Act, the GDPR and French sector-specific rules. Here is how to untangle this triple framework, for manufacturers, software vendors and user healthcare facilities alike.

First question: is your software a medical device?

Everything starts here. Software is a medical device as soon as its manufacturer assigns it a medical intended purpose: diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease. It does not matter that it does not act physically on the body: standalone software (Software as a Medical Device) is fully covered by European Regulation 2017/745 on medical devices (MDR).

The qualification criteria come down to three questions:

  • Does the software do something other than store, archive, communicate or search data? A simple electronic patient record that displays results is not a medical device; a module that interprets them may be.
  • Does it perform an action on data for the benefit of an individual patient? A departmental statistics tool is not a medical device; a score calculated for a given patient is.
  • Is its claimed purpose medical? Marketing material counts as much as the instructions for use: promising diagnostic support in a brochure is enough to claim a medical intended purpose.

On classification, MDR Rule 11 is strict for software: software intended to provide information used to take decisions for diagnostic or therapeutic purposes falls at least into class IIa, and moves up to class IIb or III if those decisions may cause serious or irreversible consequences. In practice, almost all clinical AI software is at least class IIa, which requires a notified body to be involved for CE marking: clinical evaluation, risk management, post-market surveillance.

The classic start-up trap: positioning the product as a "simple support tool" or a "wellness" product in the instructions for use while promising diagnostic performance in the sales pitch. Authorities qualify on the basis of the actual intended purpose, and requalification after placing on the market is a worst-case scenario: withdrawal, liability, sometimes a criminal offence.

The full qualification method, step by step, with seven software cases, Rule 11 and the treatment of legacy software, is set out in the test for qualifying software as a medical device.

Second layer: the AI Act

Regulation 2024/1689 on AI adds its own grid. For healthcare, the central mechanism is Article 6(1): an AI system is classified as high-risk where it constitutes a medical device (or a safety component of such a product) subject to a conformity assessment by a notified body. In other words: AI + medical device of class IIa or above = high-risk under the AI Act, almost automatically.

The resulting obligations for the manufacturer are added to those of the MDR: an AI-specific risk management system, governance of training data (representativeness, bias), technical documentation, logging, transparency towards the user, human oversight, robustness and cybersecurity. The good news: the legislator has provided for how the two fit together. The AI Act conformity assessment is integrated into the MDR one, with a single technical file possible and notified bodies as the pivot. The deadline for these embedded products is set at 2 August 2028: Regulation (EU) 2026/1744, known as the Digital Omnibus on AI, postponed it by one year. High-risk systems in Annex III follow on 2 December 2027. This postponement gives manufacturers a little breathing room, not latecomers: a medical device certification cycle is measured in years.

For user facilities, the AI Act creates the role of deployer: use in accordance with the instructions, human oversight entrusted to trained professionals, control of input data, retention of logs, information of the professionals concerned. A hospital that deploys a triage AI can no longer simply buy it: it must organise how it is used.

Third layer: the GDPR and health data

AI in healthcare feeds on health data, a special category under Article 9 of the GDPR whose processing is prohibited save for exceptions. Three moments in the life cycle concentrate the issues:

  • Training: building a health database to develop a model requires a solid legal basis (scientific research with safeguards, consent, or regulated reuse), appropriate information to individuals, genuine minimisation and, in most cases, a DPIA. Pseudonymisation is a safeguard, not a loophole: pseudonymised data remains personal data. The European Health Data Space will open an organised channel of access to this data for training, in secure environments: a major opportunity, provided the procedure is mastered.
  • Inference: in production, the software processes the patient's data to produce its output. Responsibilities between vendor and facility, the Article 28 DPA, HDS-certified hosting (French health data hosting certification) if the vendor hosts on behalf of third parties, Article 32 security: the standard foundation applies in full.
  • Continuous improvement: reusing production data to retrain the model is a new processing operation, with its own legal basis and contractual framework. It is the point most often forgotten in contracts, and the most sensitive for facilities.

The French touch: patient information and human oversight guarantee

French law adds a requirement of its own. Article L4001-3 of the French Public Health Code, introduced by the 2021 bioethics law, requires the healthcare professional who uses a medical device incorporating algorithmic data processing to inform the patient, and provides that the professional is informed of the use of this processing, that the patient data used is accessible and that the results produced are explained to them. Designers must ensure that how the system works is explainable to users.

This is the legal expression of the principle of human oversight guarantee: the algorithm informs, the professional decides. In practice, it requires vendors to design interfaces that make the system's reasoning intelligible (factors taken into account, confidence level, limits), and facilities to build patient information into care pathways. Opaque software, even CE-marked, exposes its French user to a breach.

Liability: who pays when the AI gets it wrong?

The question comes up in every negotiation. The state of the law comes down to a few markers: the medical decision remains the professional's, who answers for their faults under the usual conditions; the manufacturer answers for defects in its product, and the European product liability regime has been modernised to cover software and AI explicitly, with rules making proof easier for victims; between the parties, the contract allocates the final burden: performance warranties, caps, insurance. Three contractual reflexes: define precisely the performance promised (sensitivity, specificity, reference population), document the limits of use to frame misuse, and calibrate caps and insurance on the risk of bodily injury, not on the licence price.

Manufacturer checklist and facility checklist

You are developing a healthcare AI:

  1. Qualify the medical intended purpose and the MDR class before writing the first line of the sales brochure.
  2. Build a single MDR + AI Act technical file: risk management, training data, clinical evaluation, explainability.
  3. Secure the data chain: legal basis for training, HDS if you host, DPA, continuous improvement clause.
  4. Build the human oversight guarantee into the product: explainability, traceability, information for the professional.
  5. Treat 2 August 2028 as a regulatory go-live date, not a distant horizon.

You are deploying a healthcare AI:

  1. Require the CE marking, the class, the AI Act documentation and the complete instructions for use before any pilot.
  2. Organise human oversight: who validates, with what training, what procedure for deviations.
  3. Put performance, liabilities and the fate of production data into the contract.
  4. Update your DPIAs and patient information under Article L4001-3.
  5. Track actual use: the system's logs are also your evidence in the event of a dispute.

Three practical qualification cases

The theory makes sense on real cases, anonymised from our practice:

  • The pre-triage chatbot: a conversational assistant that asks the patient about their symptoms and directs them to the right level of care. It interprets individual data and its output influences a medical referral decision: qualification as a medical device is very hard to rule out, and the risk level depends on the error scenarios (missing a life-threatening emergency pushes it upwards). Many vendors in this segment still believe they are providing "mere information".
  • Imaging AI: detection of lesions on X-rays with a suspicion score for the radiologist. A textbook case of a medical device of class IIa at least, IIb if the consequences of a false negative are serious. AI + notified body = high-risk under the AI Act through Article 6(1): the file must be single and consistent.
  • The operating theatre scheduling optimiser: an algorithm that schedules procedures according to resources. No action on data for the benefit of an individual patient for medical purposes: outside the MDR in principle. But if it incorporates clinical criteria for prioritising patients, the line gets closer, and the GDPR applies in all cases. The nuance lies in the actual configuration, not in the brochure.

The lesson: qualification is done feature by feature, based on what the product does and what you say about it. A single piece of software may contain a medical device module and modules that are not, provided they are properly delimited in the architecture and documentation.

Clinical evaluation: what the notified body expects

CE marking of a clinical AI requires demonstrating performance and clinical benefit. In practice, the notified body expects: a precise definition of the intended purpose (population, indication, context of use, target user), validation data distinct from the training data and representative of the targeted French or European population, justified metrics (sensitivity, specificity, predictive values) with their confidence intervals, an analysis of failure cases and of populations where performance degrades, and a post-market clinical follow-up plan. The recurring stumbling block in AI files: performance demonstrated on foreign or outdated data that cannot be transposed to the population of use. Plan the building of your validation datasets as a regulatory workstream in its own right, with its own GDPR compliance.

Models that learn: the problem of updates

One feature of AI collides head-on with the logic of certification: the model evolves. Yet a substantial modification of a certified medical device in principle requires a new conformity assessment, and the AI Act takes the same approach for high-risk systems. Three operational consequences:

  • retraining a model in production is not a simple software update: it must be qualified, documented and, depending on its scale, submitted to the notified body;
  • the solution lies in predetermined change control plans: describing in advance, in the certification file, the planned changes to the model, their boundaries and the associated validation tests, so that they can be deployed without a new full procedure;
  • on the contract side, the user facility must be informed of changes to the model and their impact on performance: a version freeze clause for critical uses and a version log are now part of the standard negotiation terms.

Vendors that industrialise this version management turn a constraint into a selling point: documented control of the model's life cycle is exactly what hospital buyers have learned to demand.

The regulatory budget: what to provide for

Let us talk figures, because this is where projects go off the rails. The regulatory cost of a clinical AI combines several items: building and qualifying the training and validation datasets (often the heaviest item, between data access, annotation and GDPR compliance), the quality management system and technical documentation, clinical evaluation, notified body fees with review times measured in quarters, then post-market surveillance and version management. For a start-up, the overall order of magnitude runs into hundreds of thousands of euros and 18 to 36 months between design and CE marking of a class IIa device. Two strategic implications: the regulatory issue must appear in the financing plan presented to investors, who now systematically ask about it in due diligence; and the trade-off between developing a medical device function in-house or integrating a third party's already certified component is as much a legal and financial decision as a technical one.

Key takeaways

Healthcare AI is brought into compliance in three layers: the MDR for medical device qualification and CE marking, the AI Act for the high-risk regime with a 2 August 2028 deadline for certified products, and the GDPR for every stage of the data life cycle, all supplemented in France by patient information and the human oversight guarantee of Article L4001-3. Compliance is built product by product, in a single consistent file, and it starts with an honest qualification of the medical intended purpose.

Mirabile Avocat advises manufacturers, vendors and healthcare facilities: MDR and AI Act qualification, training data strategy, deployment contracts, liability files. Test the qualification of your software with the simulator below, then come and discuss it.

To learn more

Is AI diagnostic support software a medical device?

Almost always. As soon as the software interprets data for an individual patient with a claimed medical purpose, it falls under Regulation 2017/745. Rule 11 places it at least in class IIa where its information is used for diagnostic or therapeutic decisions, which requires a notified body for CE marking.

Is medical AI automatically high-risk under the AI Act?

In practice, yes for clinical AI: Article 6(1) of the AI Act classifies as high-risk AI systems that constitute medical devices subject to assessment by a notified body, meaning class IIa and above. The AI Act obligations are then integrated into the MDR technical file, with a deadline of 2 August 2028 for these products, following the postponement decided by the Digital Omnibus of July 2026.

What does French law say about AI used by doctors?

Article L4001-3 of the French Public Health Code requires the patient to be informed of the use of a medical device incorporating algorithmic processing, guarantees the professional access to the data used and an explanation of the results, and requires designers to make the system's operation explainable. It is the expression of the principle of human oversight guarantee: the algorithm informs, the professional decides.

Can an AI be trained on patient data?

Yes, under strict conditions: health data falls under Article 9 of the GDPR, which requires an appropriate legal basis (regulated research, consent or authorised reuse), information to individuals, minimisation and, in most cases, a DPIA. Pseudonymisation does not take the data outside the GDPR. The European Health Data Space will open an organised access channel for training, in a secure environment.

Who is liable when a medical AI makes an error?

The medical decision remains the professional's, who answers for their faults. The manufacturer answers for defects in its product, as the European product liability regime now explicitly covers software and AI. Between the parties, the contract allocates the final burden: performance promised, limits of use, caps and insurance calibrated on the risk of bodily injury. I calibrate this allocation as part of my e-health and healthtech advisory work.

Still have questions?

Our team is available!

Have a question?

Your information will remain strictly confidential.
Thank you! We will get back to you shortly. If you'd like to speed things up, schedule a time with me directly here:
Schedule a 15-minute call
Oops! Something went wrong while submitting the form.
Homme en costume bleu foncé avec cravate et pochette blanche, bras croisés, regardant vers l'avant.

Ressources

Aller plus loin

00
article(s) affiché(s) sur
00

4 min

Commercial agent: a key player in software sales
The commercial agent is a key player in software sales. In the world of digital commerce, the commercial agent plays a crucial role. They represent a company that sells software and act as the link with potential clients. This role takes on particular importance in France, where regulation and the leg

5 min

DORA Lawyer - Cybersecurity
The European regulation on digital operational resilience for the financial sector, also known as the “DORA Regulation” (Digital Operational Resilience Act), is a European Union initiative aimed at strengthening the digital resilience of financial sector players.

3 min

How to get your software development contracts right?
Are you an IT service provider wondering how to navigate the maze of software development contracts? Look no further. In this article, we break down what a software development contract is and how you can optimise its drafting to best meet your needs and those of your clients

10 min

Pre-Contractual Disclosure Document (DIP): a crucial document in your franchise project
In the world of franchising, one document stands out for its fundamental and mandatory nature: the Pre-Contractual Disclosure Document, commonly known as the DIP.

7 min

IT project failure: what remedies are available to affected businesses?
An IT project often represents substantial investment for businesses, yet many end in failure.

Artificial intelligence law

13 min

AI and Recruitment: What the AI Act and the GDPR Require of HR
CV screening, scoring, employee evaluation: HR AI tools are classed as high risk by the AI Act and are already governed by the GDPR and the French Labour Code. Obligations, discrimination risk and an 8-point deployment checklist.
Book an appointment
Book an appointment📆 15-Min Meeting